Determination of loss on ignition in alumino silicate refractory materials

 Loss on ignition in alumino silicate refractory materials

Determination of loss on ignition in alumino silicate refractory materials

Scope/Purpose:

This SOPs covers a method for determination of loss on ignition in alumino silicate refractory materials.

Principle: 

Loss on ignition of the material is determined on ignited in a muffle furnace at 1025 ± 20°C.

Apparatus/Equipment:

  • Platinum crucibles
  • Muffle Furnace (1025 ± 20°C)
  • Balance

Test Sample  

It is prepared for the purpose of chemical analysis by drying a sufficient quantity of the laboratory sample in a glass weighing bottle in air oven at 105 to 110°C.

Procedure: 

  • Weigh 1.0 g of sample into a previously weighed platinum crucible. 
  • Introduce it into a muffle furnace and slowly raise the temperature to 1025 ± 20°C. Ignite le mass at this temperature for 1 hour. 
  • Remove the crucible and cool in a desiccator containing preferably magnesium perchlorate and weigh. Repeat the heating till constant weight.

Calculations:

   Loss on ignition % = (A-B) × 100 / C 
Where,
A = initial mass in g of the crucible with sample, 
B = final mass in g of the crucible with residue after ignition, and 
C = mass in g of the sample taken.
Reference Standard: IS 12107: 1987 part -1